Abstract
A new microfluidic polymer chip with an embedded cationic surfactant (CS) ion-selective optode (CS-optode) as a detector of flow-injection analysis (FIA) for the determination of CSs was developed. The optode sensing membrane is based on a poly(vinyl chloride) membrane plasticized with 2-nitrophenyl octyl ether containing tetrabromophenolphthalein ethyl ester. Under the optimal flow conditions of the FIA system, the CS-optode showed a good linear relationship between the peak heights in the absorbance, and the concentrations of CS in a concentration range from 50 to 400 μmol dm -3 . The sample throughput of the present system for the determination of a CS ion (300 μmol dm -3 zephiramine) was ca. 11 samples h -1 . The proposed FIA system was applied to determine the level of CS in dental rinses.
Original language | English |
---|---|
Pages (from-to) | 195-199 |
Number of pages | 5 |
Journal | Analytical Sciences |
Volume | 34 |
Issue number | 2 |
DOIs | |
Publication status | Published - 2018 |
Keywords
- Cationic surfactants
- Flow injection analysis
- Microfludic polymer chip
- Optode
- Tetrabromophenolphthalein ethyl ester
ASJC Scopus subject areas
- Analytical Chemistry