A Shift‐Variant Deconvolution for Aperture Correction of Collimator in Emission CT—Experimental Verification

Seungkuen Paek, Masato Nakajima, Tatsuru Morozumi, Shin'Ichi Yuta

Research output: Contribution to journalArticlepeer-review

Abstract

One of the causes of the image‐quality degradation in emission computed‐tomography (ECT) is the characteristic of the aperture of its detection system. This is due to the geometrical structure of the collimator used for limiting the direction of the emitting ray. This kind of degrading causes a shift‐variant blurring in its reconstructed image. We have proposed a method of compensating the degradation by expanding the measured projection data into their frequency space so that the effect of the aperture is turned into a shift‐invariant blurring in each frequency [4, 5], This paper describes an application of the correction method to actual data obtained by light‐emission CT using visible light. The result confirms the usefulness of the proposed method.

Original languageEnglish
Pages (from-to)56-64
Number of pages9
JournalSystems and Computers in Japan
Volume21
Issue number2
DOIs
Publication statusPublished - 1990

ASJC Scopus subject areas

  • Theoretical Computer Science
  • Information Systems
  • Hardware and Architecture
  • Computational Theory and Mathematics

Fingerprint

Dive into the research topics of 'A Shift‐Variant Deconvolution for Aperture Correction of Collimator in Emission CT—Experimental Verification'. Together they form a unique fingerprint.

Cite this