Abstract
An amperometric thin-film glucose biosensor based on a plasma-polymerized film using hexamethyldisiloxane as the monomer is presented. The plasma-polymerized film, achieved in plasma in the vapor phase, offers a new alternative for use in the design of the electrode-enzyme interface of biosensors. The film shows promise of high sensor performance; namely, rapid sensor response, low noise, a wide dynamic range, reproducibility, and reduction in the effects of interfering materials such as ascorbic acid. In this study, we examined the usefulness of the hexamethyl-disiloxane plasma-polymerized film and investigated how the thickness of the plasma-polymerized film on a platinum electrode affected sensor characteristics: the selectivity for hydrogen peroxide versus interfering agents, the sensor response due to enzymatic reaction, and oxygen depletion.
Original language | English |
---|---|
Pages (from-to) | 142-147 |
Number of pages | 6 |
Journal | IEICE Transactions on Electronics |
Volume | E87-C |
Issue number | 2 |
Publication status | Published - 2004 Feb |
Keywords
- Biosensor
- Enzyme
- Glucose
- Hexamethyldisiloxane
- Plasma-polymerized film
ASJC Scopus subject areas
- Electronic, Optical and Magnetic Materials
- Electrical and Electronic Engineering