TY - JOUR
T1 - Compact electrostatic levitator for diffraction measurements with a two axis diffractometer and a laboratory x-ray source
AU - Masaki, Tadahiko
AU - Ishikawa, Takehiko
AU - Paradis, Paul Fraņois
AU - Yoda, Shinichi
AU - Okada, Junpei T.
AU - Watanabe, Yasuhiro
AU - Nanao, Susumu
AU - Ishikura, Akiko
AU - Higuchi, Kensuke
AU - Mizuno, Akitoshi
AU - Watanabe, Masato
AU - Kohara, Shinji
PY - 2007
Y1 - 2007
N2 - A compact electrostatic levitator was developed for the structural analysis of high-temperature liquids by x-ray diffraction methods. The size of the levitator was 200 mm in diameter and 200 mm in height and can be set up on a two axis diffractometer with a laboratory x-ray source, which is very convenient in performing structural measurements of high-temperature liquids. In particular, since the laboratory x-ray source allows a great amount of user time, preliminary or challenging experiments can be performed with trial and error, which prepares and complements synchrotron x-ray experiments. The present small apparatus also provides the advantage of portability and facility of setting. To demonstrate the capability of this electrostatic levitator, the static structure factors of alumina and silicon samples in their liquid phases were successfully measured.
AB - A compact electrostatic levitator was developed for the structural analysis of high-temperature liquids by x-ray diffraction methods. The size of the levitator was 200 mm in diameter and 200 mm in height and can be set up on a two axis diffractometer with a laboratory x-ray source, which is very convenient in performing structural measurements of high-temperature liquids. In particular, since the laboratory x-ray source allows a great amount of user time, preliminary or challenging experiments can be performed with trial and error, which prepares and complements synchrotron x-ray experiments. The present small apparatus also provides the advantage of portability and facility of setting. To demonstrate the capability of this electrostatic levitator, the static structure factors of alumina and silicon samples in their liquid phases were successfully measured.
UR - http://www.scopus.com/inward/record.url?scp=33847654205&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=33847654205&partnerID=8YFLogxK
U2 - 10.1063/1.2435590
DO - 10.1063/1.2435590
M3 - Article
AN - SCOPUS:33847654205
SN - 0034-6748
VL - 78
JO - Review of Scientific Instruments
JF - Review of Scientific Instruments
IS - 2
M1 - 026102