Relationship of Catheter Contact Angle and Contact Force with Contact Area on the Surface of Heart Muscle Tissue in Cardiac Catheter Ablation

Kriengsak Masnok, Nobuo Watanabe

Research output: Contribution to journalArticlepeer-review

9 Citations (Scopus)

Abstract

Purpose: The aims of this study were to develop an experimental procedure for setting the catheter angle with respect to the surface of the heart muscle and the catheter contact force and to investigate the catheter contact area on the heart muscle as a function of catheter contact angle and force. Methods: Visualization tests were performed for 5 contact angles (0°, 30°, 45°, 60°, and 90°) and 8 contact forces (2, 4, 6, 10, 15, 20, 30, and 40 gf). Each experiment was repeated 6 times with 2 different commercially available catheter tips. Results: The morphology of the contact area was classified into rectangular, circular, ellipsoidal, and semi-ellipsoidal. The correlation between contact force and contact area was a logarithmic function; increasing contact force was associated with increased contact area. At the same contact force, the correlation between contact angle and contact area was inverse; decreasing contact angle was associated with a corresponding increase in contact area. Conclusion: Both the catheter contact angle and contact force substantially impact the contact area and morphology in catheter ablation procedures.

Original languageEnglish
Pages (from-to)407-417
Number of pages11
JournalCardiovascular Engineering and Technology
Volume12
Issue number4
DOIs
Publication statusPublished - 2021 Aug

Keywords

  • Catheter ablation
  • Contact angle
  • Contact area
  • Contact force
  • Tachyarrhythmia

ASJC Scopus subject areas

  • Biomedical Engineering
  • Cardiology and Cardiovascular Medicine

Fingerprint

Dive into the research topics of 'Relationship of Catheter Contact Angle and Contact Force with Contact Area on the Surface of Heart Muscle Tissue in Cardiac Catheter Ablation'. Together they form a unique fingerprint.

Cite this