TY - JOUR
T1 - Sequential injection chemiluminescence immunoassay for anionic surfactants using magnetic microbeads immobilized with an antibody
AU - Zhang, Ruiqi
AU - Hirakawa, Koji
AU - Seto, Daisuke
AU - Soh, Nobuaki
AU - Nakano, Koji
AU - Masadome, Takashi
AU - Nagata, Kazumi
AU - Sakamoto, Kazuhira
AU - Imato, Toshihiko
N1 - Funding Information:
The authors are grateful for financial support from Foundation of Fukuoka Coal Mining Area Development. The authors also wish to thank Dr. Tomohiro Uchimura (Kyushu University) for his helpful advice on the arrangement of optical devices.
PY - 2005/12/15
Y1 - 2005/12/15
N2 - A rapid and sensitive immunoassay for the determination of linear alkylbenzene sulfonates (LAS) is described. The method involves a sequential injection analysis (SIA) system equipped with a chemiluminescence detector and a neodymium magnet. Magnetic beads, to which an anti-LAS monoclonal antibody was immobilized, were used as a solid support in an immunoassay. The introduction, trapping and release of the magnetic beads in the flow cell were controlled by means of a neodymium magnet and adjusting the flow of the carrier solution. The immunoassay was based on an indirect competitive immunoreaction of an anti-LAS monoclonal antibody on the magnetic beads and the LAS sample and horseradish peroxidase (HRP)-labeled LAS, and was based on the subsequent chemiluminscence reaction of HRP with hydrogen peroxide and p-iodophenol, in a luminol solution. The anti-LAS antibody was immobilized on the beads by coupling the antibody with the magnetic beads after activation of a carboxylate moiety on the surface of magnetic beads that had been coated with a polylactic acid film. The antibody immobilized magnetic beads were introduced, and trapped in the flow cell equipped with the neodymium magnet, an LAS solution containing HRP-labeled LAS at constant concentration and the luminol solution were sequentially introduced into the flow cell based on an SIA programmed sequence. Chemiluminescence emission was monitored by means of a photon counting unit located at the upper side of the flow cell by collecting the emitted light with a lens. A typical sigmoid calibration curve was obtained, when the logarithm of the concentration of LAS was plotted against the chemiluminescence intensity using various concentrations of standard LAS samples (0-500 ppb) under optimum conditions. The time required for analysis is less than 15 min.
AB - A rapid and sensitive immunoassay for the determination of linear alkylbenzene sulfonates (LAS) is described. The method involves a sequential injection analysis (SIA) system equipped with a chemiluminescence detector and a neodymium magnet. Magnetic beads, to which an anti-LAS monoclonal antibody was immobilized, were used as a solid support in an immunoassay. The introduction, trapping and release of the magnetic beads in the flow cell were controlled by means of a neodymium magnet and adjusting the flow of the carrier solution. The immunoassay was based on an indirect competitive immunoreaction of an anti-LAS monoclonal antibody on the magnetic beads and the LAS sample and horseradish peroxidase (HRP)-labeled LAS, and was based on the subsequent chemiluminscence reaction of HRP with hydrogen peroxide and p-iodophenol, in a luminol solution. The anti-LAS antibody was immobilized on the beads by coupling the antibody with the magnetic beads after activation of a carboxylate moiety on the surface of magnetic beads that had been coated with a polylactic acid film. The antibody immobilized magnetic beads were introduced, and trapped in the flow cell equipped with the neodymium magnet, an LAS solution containing HRP-labeled LAS at constant concentration and the luminol solution were sequentially introduced into the flow cell based on an SIA programmed sequence. Chemiluminescence emission was monitored by means of a photon counting unit located at the upper side of the flow cell by collecting the emitted light with a lens. A typical sigmoid calibration curve was obtained, when the logarithm of the concentration of LAS was plotted against the chemiluminescence intensity using various concentrations of standard LAS samples (0-500 ppb) under optimum conditions. The time required for analysis is less than 15 min.
KW - Anionic surfactants
KW - Chemiluminescence
KW - Immunoassay
KW - Magnetic microbeads
KW - Sequential injection
UR - http://www.scopus.com/inward/record.url?scp=27744594319&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=27744594319&partnerID=8YFLogxK
U2 - 10.1016/j.talanta.2005.07.012
DO - 10.1016/j.talanta.2005.07.012
M3 - Article
C2 - 18970310
AN - SCOPUS:27744594319
SN - 0039-9140
VL - 68
SP - 231
EP - 238
JO - Talanta
JF - Talanta
IS - 2
ER -