Abstract
Purpose: The purpose of this study is to improve the calcification depiction ability in ultrasonography using correlation of echoes with a modified Wiener filter. Methods: The waveform of an ultrasound pulse changes when it passes through the location of a calcification. Since the change in echo waveform caused by a calcification decreases the correlation of waveforms in adjacent scan lines, we have proposed a calcification depiction method using the decorrelation of echoes. However, the low signal-to-noise ratio of echoes also decreases the correlation of the echoes. In this study, we employ the correlation of echoes with a modified Wiener filter to suppress the effect of noise, as an indicator of a calcification. Results: The proposed calcification indicator depicted copper cylindrical rods 0.2 mm in size at a depth of 2 cm with a sensitivity of 80% and a positive predictive value of 80%, despite being hardly depicted at all on B-mode ultrasound imaging. Conclusion: This study suggests the potential of the proposed method to improve the performance of calcification depiction by ultrasound devices.
Original language | English |
---|---|
Pages (from-to) | 127-135 |
Number of pages | 9 |
Journal | Journal of Medical Ultrasonics |
Volume | 39 |
Issue number | 3 |
DOIs | |
Publication status | Published - 2012 Jul |
Externally published | Yes |
Keywords
- Calcification
- Calcification depiction
- Decorrelation
- Ultrasonography
- Wiener filter