抄録
To monitor the changes in the concentration of antibody produced in body fluids due to infection, we developed a new method for continuous measuring of the antibody concentration. Conventional immunoassay is applicable to heterogeneous reaction between antigen and antibody, and is unavailable for continuous measurement. We used the technique of electrochemiluminescence (ECL) to measure antibody concentration continuously in a homogeneous reaction system. We used human serum albumin (HSA) as antigen and luminol as luminous substance. ECL reaction occurred on electrically conductive glass electrode the electric potential of which was kept constant at -0.70 V vs saturated calomel electrode (SCE). In conclusion, antibody concentration can be measured continuously by the ECL reaction which is a promising technique of measurement of antibody specifically reacting with antigen.
本文言語 | English |
---|---|
ページ(範囲) | 523-526 |
ページ数 | 4 |
ジャーナル | Japanese Journal of Artificial Organs |
巻 | 23 |
号 | 2 |
出版ステータス | Published - 1994 1月 1 |
外部発表 | はい |
ASJC Scopus subject areas
- 生物理学