抄録
A flow injection system for the determination of nonionic surfactants in the presence of ionic surfactants was investigated. The system consists of a microcolumn packed with an ion-exchange resin and a surfactant-selective electrode using a plasticized polyvinyl chloride membrane containing no ion-exchange site. Ionic surfactants were removed on the microcolumn. Nonionic surfactants passed through the column with ethanol/H20 = 50/50 (v/v, %) were detected. The electrode showed the cationic potential response with a slope of 40 mV/decade for Triton X-100 from 4×10-5 to 5×10-4 M. The detection limit of Triton X-100 was 1 × 10-5 M. The proposed method is free from the interference of coexisting inorganic electrolytes in sample solutions. By this method, nonionic surfactants in detergent samples could be determined.
本文言語 | English |
---|---|
ページ(範囲) | 7-11 |
ページ数 | 5 |
ジャーナル | Bunseki Kagaku |
巻 | 40 |
号 | 1 |
DOI | |
出版ステータス | Published - 1991 1月 |
外部発表 | はい |
ASJC Scopus subject areas
- 分析化学