TY - JOUR
T1 - Fluorometric gas-imaging system (sniff-cam), using the extinction of NADH with an ADH reverse reaction, for acetaldehyde in the gas phase
AU - Iitani, Kenta
AU - Sato, Toshiyuki
AU - Naisierding, Munire
AU - Hayakawa, Yuuki
AU - Toma, Koji
AU - Arakawa, Takahiro
AU - Mitsubayashi, Kohji
N1 - Funding Information:
This work was supported by JSPS KAKENHI Grant Numbers JP17H01759, JP16J09604, the Japan Science and Technology Agency (JST), and the Ministry of Education, Culture, Sports, Science and Technology (MEXT) Special Funds for Education and Research “Advanced Research Program in Neo-Biology”.
Publisher Copyright:
© 2017 The Royal Society of Chemistry.
PY - 2017/10/21
Y1 - 2017/10/21
N2 - A gas-imaging system (sniff-cam) that allows fluorometric visualization of a two-dimensional (2-D) distribution of gaseous acetaldehyde (AcH) was developed. It employed a reverse reaction of a nicotinamide adenine dinucleotide (NADH) dependent enzyme that led to consumption of NADH in that reaction. The system was constructed with a highly sensitive camera, an ultraviolet light emitting diode array sheet, two band pass filters and an alcohol dehydrogenase (ADH)-immobilized mesh that was used for AcH detection. The reverse reaction of the ADH catalyzed the reduction of AcH to ethanol and the oxidation of NADH to NAD+, which occurred when gaseous AcH was applied to the ADH immobilized mesh that was wetted with a slightly acidic NADH solution. As NADH has an autofluorescence property [emission (λem) at 490 nm; excitation (λex) at 340 nm], the presence of gaseous AcH was visualized by a decrease of fluorescence of the NADH at the ADH immobilized mesh. After constructing the gaseous AcH imaging system, optimizations of pH, and concentration of the NADH solution were performed. As a result of the optimizations (500 μM of NADH in 0.1 M of Tris hydrochloride (Tris-HCl) buffer at pH 6.5), the AcH sniff-cam showed a wide dynamic range (0.1-10 ppm) for gaseous AcH with a high correlation coefficient (R = 0.999). Furthermore, a fluorescence gradient with a rounded shape centered in a gas outlet was observed. These results demonstrated that the AcH sniff-cam utilizing the fluorescence decrease of NADH could be used to quantitatively evaluate the 2-D distribution of gaseous AcH.
AB - A gas-imaging system (sniff-cam) that allows fluorometric visualization of a two-dimensional (2-D) distribution of gaseous acetaldehyde (AcH) was developed. It employed a reverse reaction of a nicotinamide adenine dinucleotide (NADH) dependent enzyme that led to consumption of NADH in that reaction. The system was constructed with a highly sensitive camera, an ultraviolet light emitting diode array sheet, two band pass filters and an alcohol dehydrogenase (ADH)-immobilized mesh that was used for AcH detection. The reverse reaction of the ADH catalyzed the reduction of AcH to ethanol and the oxidation of NADH to NAD+, which occurred when gaseous AcH was applied to the ADH immobilized mesh that was wetted with a slightly acidic NADH solution. As NADH has an autofluorescence property [emission (λem) at 490 nm; excitation (λex) at 340 nm], the presence of gaseous AcH was visualized by a decrease of fluorescence of the NADH at the ADH immobilized mesh. After constructing the gaseous AcH imaging system, optimizations of pH, and concentration of the NADH solution were performed. As a result of the optimizations (500 μM of NADH in 0.1 M of Tris hydrochloride (Tris-HCl) buffer at pH 6.5), the AcH sniff-cam showed a wide dynamic range (0.1-10 ppm) for gaseous AcH with a high correlation coefficient (R = 0.999). Furthermore, a fluorescence gradient with a rounded shape centered in a gas outlet was observed. These results demonstrated that the AcH sniff-cam utilizing the fluorescence decrease of NADH could be used to quantitatively evaluate the 2-D distribution of gaseous AcH.
UR - http://www.scopus.com/inward/record.url?scp=85030979264&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=85030979264&partnerID=8YFLogxK
U2 - 10.1039/c7an00524e
DO - 10.1039/c7an00524e
M3 - Article
C2 - 28758645
AN - SCOPUS:85030979264
SN - 0003-2654
VL - 142
SP - 3830
EP - 3836
JO - The Analyst
JF - The Analyst
IS - 20
ER -