TY - JOUR
T1 - Integration of microfabricated needle-type glucose sensor devices with a novel thin-film Ag/AgCl electrode and plasma-polymerized thin film
T2 - Mass production techniques
AU - Hiratsuka, Atsunori
AU - Kojima, Ken ichi
AU - Suzuki, Hiroaki
AU - Muguruma, Hitoshi
AU - Ikebukuro, Kazunori
AU - Karube, Isao
N1 - Copyright:
Copyright 2020 Elsevier B.V., All rights reserved.
PY - 2001
Y1 - 2001
N2 - We developed an integrated array of needle-type biosensors employing a novel process of fabrication, comprising conventional semiconductor fabrication and micromachining technology. Amperometric sensing electrodes with plasma-polymerized films and a thin-film Ag/AgCl reference electrode were directly integrated on a glass substrate with thin-film process, e.g., sputtering. An enzyme was immobilized on the electrode via the plasma-polymerized film, which was deposited directly on the substrate using a dry process. The novel thin-film Ag/AgCl reference electrode showed stable potentials in concentrated chloride solutions for a long period. The plasma-polymerized film is considered to play an important role as an interfacial design between the sensing electrode and the immobilized enzyme considering that the film is extremely thin, adheres well to the substrate (electrode) and has a highly cross-linked network structure and functional groups, such as amino groups. The results showed increments of the sensor signal, probably because the plasma-polymerized film allowed a large amount of enzyme to be immobilized. The greatest advantage is that the process can permit the mass production of high-quality biosensors at a low cost.
AB - We developed an integrated array of needle-type biosensors employing a novel process of fabrication, comprising conventional semiconductor fabrication and micromachining technology. Amperometric sensing electrodes with plasma-polymerized films and a thin-film Ag/AgCl reference electrode were directly integrated on a glass substrate with thin-film process, e.g., sputtering. An enzyme was immobilized on the electrode via the plasma-polymerized film, which was deposited directly on the substrate using a dry process. The novel thin-film Ag/AgCl reference electrode showed stable potentials in concentrated chloride solutions for a long period. The plasma-polymerized film is considered to play an important role as an interfacial design between the sensing electrode and the immobilized enzyme considering that the film is extremely thin, adheres well to the substrate (electrode) and has a highly cross-linked network structure and functional groups, such as amino groups. The results showed increments of the sensor signal, probably because the plasma-polymerized film allowed a large amount of enzyme to be immobilized. The greatest advantage is that the process can permit the mass production of high-quality biosensors at a low cost.
UR - http://www.scopus.com/inward/record.url?scp=0142121113&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=0142121113&partnerID=8YFLogxK
U2 - 10.1039/b008036p
DO - 10.1039/b008036p
M3 - Article
C2 - 11394309
AN - SCOPUS:0142121113
SN - 0003-2654
VL - 126
SP - 658
EP - 663
JO - The Analyst
JF - The Analyst
IS - 5
ER -