T1 - Microfluidic polymer chip with an embedded ion-selective electrode detector for nitrate-ion assay in environmental samples
AU - Masadome, Takashi
AU - Nakamura, Kazuki
AU - Iijima, Daisuke
AU - Horiuchi, Osamu
AU - Tossanaitada, Benjaporn
AU - Wakida, Shin ichi
AU - Imato, Toshihiko
PY - 2010/4
Y1 - 2010/4
N2 - A nitrate ion-selective electrode (NO3--ISE) has been developed based on tetradodecylammonium bromide as an anion exchanger and 2-nitrophenyl octyl ether as a plasticizer. The NO3--ISE shows an almost Nernstian response to nitrate ion over a concentration range between 1.0 × 10-6 and 1.0 × 10-1 M, with an anionic slope of -57.7 ± 0.7 mV/decade. The selectivity coefficients of the NO3--ISE for nitrate ion against chloride and sulfate were obtained. A microfluidic polymer chip was fabricated using polystyrene plates and stainless-steel wires as a template for the channel. The microfluidic polymer chip is composed of a mixing chip and a NO3--ISE detector chip. The microfluidic polymer chip, integrated with a NO3--ISE detector consisting of the NO3--ISE and a Na+-ISE as a reference electrode, showed an almost Nernstian response to nitrate ion over a concentration range between 1.0 × 10-5 and 1.0 × 10-1 M, with an anionic slope of -54.3 ± 1.3 mV/decade. The microfluidic polymer chip was then applied to the determination of nitrate ion in environmental water samples, such as a tap water, a well-water sample and water for agricultural use. 2010
AB - A nitrate ion-selective electrode (NO3--ISE) has been developed based on tetradodecylammonium bromide as an anion exchanger and 2-nitrophenyl octyl ether as a plasticizer. The NO3--ISE shows an almost Nernstian response to nitrate ion over a concentration range between 1.0 × 10-6 and 1.0 × 10-1 M, with an anionic slope of -57.7 ± 0.7 mV/decade. The selectivity coefficients of the NO3--ISE for nitrate ion against chloride and sulfate were obtained. A microfluidic polymer chip was fabricated using polystyrene plates and stainless-steel wires as a template for the channel. The microfluidic polymer chip is composed of a mixing chip and a NO3--ISE detector chip. The microfluidic polymer chip, integrated with a NO3--ISE detector consisting of the NO3--ISE and a Na+-ISE as a reference electrode, showed an almost Nernstian response to nitrate ion over a concentration range between 1.0 × 10-5 and 1.0 × 10-1 M, with an anionic slope of -54.3 ± 1.3 mV/decade. The microfluidic polymer chip was then applied to the determination of nitrate ion in environmental water samples, such as a tap water, a well-water sample and water for agricultural use. 2010
UR - http://www.scopus.com/inward/record.url?scp=77953583240&partnerID=8YFLogxK
UR - http://www.scopus.com/inward/citedby.url?scp=77953583240&partnerID=8YFLogxK
U2 - 10.2116/analsci.26.417
DO - 10.2116/analsci.26.417
M3 - Article
C2 - 20410562
AN - SCOPUS:77953583240
SN - 0910-6340
VL - 26
SP - 417
EP - 423
JO - Analytical Sciences
JF - Analytical Sciences
IS - 4
ER -