Simulation of the BP-80 blood pump

Nobuo Watanabe, Oezcan Karsak, Frank Neudel, Thomas Kink, Joern Apel, Tetsuo Fujimoto, Helmut Reul, Setsuo Takatani

研究成果: Article査読

6 被引用数 (Scopus)

抄録

In this study, computational fluid dynamics (CFD) analysis was applied to investigate the flow within a commercially available Biopump, BP-80 (Medtronic, Minneapolis, MN, U.S.A.). The Biopump was selected because, for this purpose, a great number of experimental hemolysis data is available. The process of geometry representation and grid generation was focused on, due toits high impact on the numerical results. This process incorporated the use of three commercially available software packages for three-dimensional computer-aided design (3D-CAD), grid generation, and solving, respectively. For the purpose of validation, the head/flow characteristics of the pump were experimentally obtained and compared to the computed data. The results showed a rough agreement between CFD data and experimental data. Further investigations should cover detailed shear stress analyses and computation of other hemolysis-related quantities.

本文言語English
ページ(範囲)733-739
ページ数7
ジャーナルArtificial Organs
25
9
DOI
出版ステータスPublished - 2001

ASJC Scopus subject areas

  • バイオエンジニアリング
  • 医学(その他)
  • 生体材料
  • 生体医工学

フィンガープリント

「Simulation of the BP-80 blood pump」の研究トピックを掘り下げます。これらがまとまってユニークなフィンガープリントを構成します。

引用スタイル